Overview
Specs

THE TUTTNAUER 2340EA IS A FULLY AUTOMATIC TABLETOP AUTOCLAVE DESIGNED FOR THE STERILIZATION OF WRAPPED AND UNWRAPPED INSTRUMENTS, LIQUIDS, AND LABORATORY EQUIPMENT. IT FEATURES A 19-LITER (5-GALLON) STERILIZATION CHAMBER CONSTRUCTED FROM CORROSION-RESISTANT 316L GRADE STAINLESS STEEL, MEASURING 9 INCHES IN DIAMETER BY 18 INCHES IN DEPTH. THE UNIT INCLUDES THREE STAINLESS STEEL TRAYS AND HAS A CAPACITY TO ACCOMMODATE TWO FULL AND TWO HALF-SIZE STANDARD CASSETTES. THIS MODEL IS EQUIPPED WITH A HIGH-EFFICIENCY AIR PUMP THAT FACILITATES CLOSED-DOOR ACTIVE DRYING BY CIRCULATING HEATED AIR THROUGH A 0.2 ΜM HEPA FILTER, ENSURING INSTRUMENTS ARE THOROUGHLY DRIED WHILE MAINTAINING STERILITY. OPERATING ON A MICROPROCESSOR CONTROL SYSTEM WITH A DIGITAL DISPLAY FOR TEMPERATURE AND PRESSURE, THE 2340EA OFFERS THREE PRESET STERILIZATION PROGRAMS AT TEMPERATURES UP TO 134°C (273°F) AND ONE DRY-ONLY PROGRAM. LABORATORY PERSONNEL CAN CUSTOMIZE CYCLE PARAMETERS, INCLUDING TEMPERATURE (105 TO 137°C), STERILIZATION TIME (0 TO 59 MINUTES), AND DRYING TIME (0 TO 99 MINUTES). SAFETY FEATURES INCLUDE A DOUBLE-LOCKING DEVICE THAT PREVENTS THE DOOR FROM OPENING UNDER PRESSURE, AN AUTOMATIC SAFETY SHUTOFF TO PREVENT OVERHEATING, AND A LOW-WATER SENSOR TO PROTECT HEATING ELEMENTS. TECHNICAL SPECIFICATIONS INCLUDE A POWER RATING OF 1400W AT 120V (50/60 HZ), OVERALL DIMENSIONS OF 21.5" D X 20" W X 14.4" H, AND A WEIGHT OF APPROXIMATELY 79 LBS. THE UNIT COMPLIES WITH INTERNATIONAL STANDARDS, INCLUDING ASME, UL, AND FDA REGULATIONS.